AllergyScreen® and AlleisaScreen® are two test systems to detect allergen-specific IgE in human serum/plasma. Customers can be provided with the necessary test strips and the corresponding hard- and software (scanner, testing equipment).

Statistical data of the AllergyScreen® / AlleisaScreen® test

Statistical data:
variation, sensitivity, specificity, accuracy, linearity, LoB and LoD (based on decimal classes) - most tested allergen groups are considered

Variation

Sample level (classes) Coefficients of variation (%)
AllergyScreen® AlleisaScreen®
Intra-assay variation Inter-assay variation Intra-assay variation Inter-assay variation
Class 1 (cut-off) 27.0 26.4 25.8 18.1
Class 2-6 3.1 4.5 2.5 2.5

Inter-batch variation: mean value 8.2 % (AllergyScreen®/AlleisaScreen®)

Analytical performance of most tested allergens with an ELISA reference test (4820 determinations)

The sensitivity, specificity, concordance with the respective confidence intervals of most tested allergens over seven studies with a pairwise ELISA reference test in general (4820 determinations) are given in table below:

Sensitivity 95% confidence interval Specificity 95% confidence interval Concordance 95% confidence interval
85.6% 84.3% to 86.8% 92.4% 91.1% to 93.5% 88.3% 87.4% to 89.2%

Linearity

Linearity between the measurement signal and the concentration of the analyte is, if r is > 0.97 and is > 0.95.

As seen in the table below, all calculated regression coefficients (r and R²) are close to 1 thus demonstrating a clear positive linear relationship between the measuring signal of the AllergyScreen®/AlleisaScreen® and the allergen concentration in the sample. Although the AllergyScreen®/AlleisaScreen® is determined as a semi-quantitative assay, the linearity is very good over a broad range of dilutions as it is recommended in quantitative assays.

AllergyScreen®/AlleisaScreen® Regression coefficient r Coefficient of determination R²
Mean AllergyScreen® 0.994 0.987
Mean AlleisaScreen® 0.993 0.987

LoD and LoB

The limit of detection (LoD) was determined by the measurement of samples known to contain a low concentration of analyte and the limit of blank (LoB) by measuring a high number (>100) of blank samples. The following tables show that the LoD and the LoB of the AllergyScreen®/AlleisaScreen® are in the range of equivalent devices and the mathematical cut-off of ≥ 0.35 IU/ml has compiled with.

LoB: Meanblank + 1.645 (SDblank)

Parameter AllergyScreen® AlleisaScreen®
Blanks 402 626
Meanblank 0.01 IU/ml 0.01 IU/ml
SDblank 0.03 0.03
LoB 0.06 IU/ml 0.05 IU/ml

LoD: LoB + 1.645 (SD low concentration sample)

Parameter AllergyScreen® AlleisaScreen®
Low concentration 83 123
Meanblank 0.39 IU/ml 0.35 IU/ml
SDblank 0.17 0.16
LoD 0.34 IU/ml 0.31 IU/ml

Furthermore, an interference study showed that there is no indication of a significant influence of different blood components (bilirubin, haemoglobin, lipids) on the results of the AllergyScreen®/AlleisaScreen® test. Only a very high IgE-level in patient's serum/plasma might lead to a minor to moderate reduction of the measured values in some cases due to the hook effect. But however, this does not lead to false-negative results. Patients treated with Omalizumab and similar drugs cannot be analysed with this system (Mizuma et al. 2015).